Digitizing Resolution of the Leica GR10 Receiver


 

While differencing the L2 and L2C carrier phase measurements made with a Lecia GR10 receiver, we have observed a banding pattern; the difference between each band is ~0.002047434 cycles.

Using the L2 wavelength we found that 0.002047434 cycles = 0.5 mm

This banding demonstrates that the digitizing resolution of the GR10 receiver is 0.5 mm for the L2 signal.

This analysis was done using a translation of m00-format data using extended resolution on the phase values beyond the usual 0.001 cycle resolution available in RINEX observation format.

The following figure shows the difference between the L2 and L2C carrier phase measurements for PRN 7 from a GR10 receiver.



Article ID: 692
Created: February 3, 2011
Last Updated: August 12, 2016
Author: Henry Berglund