There has been a high number of met pack failures.
The Vaisala engineers determined that the cause is due to moisture getting into the service port.

If you recall, if you are plugged into the service port the serial connect is shut down.
So you always have to remember to unplug the USB cable to test the unit.

The moisture was shorting out this port and thus shutting down the serial port.

They have come up with a new more robust connector and will be installing it on future units.

vaisala_connector.jpg

Attached Files (1)
Comments
There are no comments for this article. Be the first to post a comment.
Name
Email
Security Code Security Code
Related Articles RSS Feed
Vaisala Weather Transmitter WXT510 Product Brochure and Specification Sheet
Viewed 2080 times since Wed, Feb 18, 2009
Vaisala - How to integrate the Vaisala WXT520 with the Topcon NET-G3
Viewed 7231 times since Tue, Jan 26, 2010
How to integrate the Vaisala WXT510/520 and the Septentrio PolaRx5 (Command Line)
Viewed 2586 times since Fri, Jan 8, 2016
How to integrate the Vaisala WXT510/520 and the Septentrio PolaRx5 (Web User Interface)
Viewed 2325 times since Tue, Apr 19, 2016
How to Mount a Vaisala WXT 520 Met Pack
Viewed 6210 times since Mon, Dec 20, 2010
WXT520 Ordering guide
Viewed 2139 times since Mon, Mar 24, 2014
Vaisala - How to integrate the Vaisala WXT520 with the Ashtech MicroZ
Viewed 5591 times since Tue, Jan 26, 2010
Vaisala Weather Transmitter WXT510 User's Guide (July 2005)
Viewed 1670 times since Fri, Nov 20, 2009
Vaisala Weather Transmitter WXT510 Order Guide (USD)
Viewed 1149 times since Wed, Feb 18, 2009
Vaisala Weather Transmitter WXT520 User's Guide (2008)
Viewed 1356 times since Mon, Feb 1, 2010
MENU

Last modified: Monday, 12-May-2014 00:18:29 UTC